Olympus Winter & Ibe GmbH Telescope, 10 mm, 30°, HD, quick lock, autoclavable; Rigid optical laparoscope
|
Back to Search Results |
|
| Model Number |
WA53005A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the rigid optical laparoscope to the service center for a separate issue.During device analysis, the service repair technician identified damage to the tip cover glass, with resulting glass fragments causing reflections in the image.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of malfunction identified during the device evaluation was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|