| Catalog Number |
VC10A256F60 |
| Medical Device Problem Codes |
Contamination (1120); Material Deformation (2976); Excessive Heating (4030)
|
| Health Effect - Clinical Codes |
Pain (1994); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a radiofrequency ablation procedure using the venclose catheter.During the procedure, the catheter allegedly heated above the temperature displayed on the generator, and a thick clot formed on the tip of the catheter that could not be removed as expected.Additionally, patient experienced pain and presented to the emergency department; ultrasound showed no ehitt or dvt.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6)2026, a patient underwent a radiofrequency ablation procedure using the venclose catheter.During the procedure, the catheter allegedly heated above the temperature displayed on the generator, and a thick clot formed on the tip of the catheter that could not be removed as expected.Additionally, patient experienced pain and presented to the emergency department; ultrasound showed no ehitt or dvt.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a radiofrequency ablation procedure using the venclose catheter.During the procedure, the catheter allegedly heated above the temperature displayed on the generator, and a thick clot formed on the tip of the catheter that could not be removed as expected.Additionally, it was reported that the patient experienced pain and presented to the emergency department (ed).The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional information was received, and the file was reassessed for reportability.Based on the medical affairs team assessment and the follow up information received, the event does not meet the criteria for a serious injury.Follow up documentation notes no ehitt or dvt identified on ultrasound, no reported medical intervention beyond evaluation, and no evidence of lasting injury, permanent impairment, or damage to a body structure or function.Therefore, the complaint was downgraded from serious injury to not reportable.Although the event is now considered non reportable, an initial mdr had already been submitted based on the information available at the time of reporting.However, following review of the additional information received, the final reportability assessment has been updated to not reportable.B5, e1, g3, h6 (patient).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr and supplemental report 1 listed an incorrect manufacturing location (g1).The correct manufacturing location for this device is viant costa rica, s.A.Additional information was received, and the file was reassessed for reportability.As a blood clot was identified on the distal tip of the catheter, the event should have been documented as a contamination issue.Therefore, the complaint was reassessed and determined to be reportable as a malfunction.B1, g1, g3, h6 (device).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|