| Model Number |
IntelliVue Multi Measurement Server SLCP |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device is not reading ecg values correctly.The device was in use on a patient at the time of the event.There was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips remote service engineer spoke to the customer biomed who stated that the ecg is not providing the correct values.The biomed was requesting parts identification for a replacement ecg-5ld/masimo spo2/w pt nrohs pca board.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was faulty ecg-5ld/masimo spo2/w pt nrohs pca board.The biomed was provide parts identification and a quote for a replacement pca ecg-5ld/masimo spo2/w pt nrohs board.The customer is taking responsibility for servicing the device.
|
| |
|