| Model Number |
861288 |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on the heartstart mrx monitor/defibrillator indicating the device can only be discharged by pressing the button, but not via the pads causing failure of the functional verification test.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is based on information provided by philips remote service engineer (rse) and has been investigated by the philips complaint handling team.The rse determined that this model is out of support, and the spare parts are no longer available due to end of life.The customer will be informed that service can no longer be completed on the unit as the device has reached end of life (eol).The heartstart mrx device and all associated service/support was discontinued on (b)(6) 2022.The device remains at the customer site, requiring no further evaluation.Based on the information available and the testing conducted, the cause of the reported problem was not determined.The customer will be informed that service can no longer be completed on the unit as the device has reached end of life (eol).Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received the complaint file will be reopened.
|
| |
|