| Model Number |
2110 |
| Medical Device Problem Code |
Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when putting the cassette, the pump presents the following message: not attached properly or reattach cassette.Pump alarms only when no cassette is attached, and the latch is closed.This is because the fluid ingress causes sensor drift; dso counts are higher than the software expects with no cassette attached, and therefore the pump alarms ¿cassette not attached properly¿.Dso might need calibration.There was no patient involvement or harm.
|
| |
|
Additional Manufacturer Narrative
|
|
No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filled as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|