| Model Number |
IntelliVue Patient Monitor MX550 |
| Medical Device Problem Code |
No Audible Alarm (1019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on an intellivue mx550 patient monitor indicating there was a speaker malfunction.It was confirmed there was no audible sound produced.The device was not in use on patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by remote service personnel which included remote communication and troubleshooting with the customer biomedical engineer.The results of the analysis indicate the device was displaying a speaker malfunction error and there was no sound produced.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a faulty mainboard.The customer ordered a replacement mainboard to resolve the issue.A review of the service history for this device shows the issue has not been reported again for this device.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|
| |
|