| Model Number |
Avalon Toco+ Transducer |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on an avalon toco transducer indicating that at the first measurement the sensor freezes on the value and does not move.The device was in use on a patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by remote service personnel which included remote communication/trouble shooting with the customer.The results of the analysis indicated the transducer was not working properly.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a faulty transducer.The issue was resolved by replacing the transducer.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|
| |
|