Catalog Number 11-300820
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Loss of Range of Motion (2032)
Date of Event 03/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately two months post-implantation, the patient underwent a hip revision due to the patient being unable to internally rotate their foot.There was a suspected failure of the taper junction or proximal body locking screw.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).H6: proposed component code: mechanical (g04) ¿ stem.Complaint is not confirmed.Visual examination of the provided picture identified explanted stem and screw, bio-debris present, however it cannot be used to confirm the reported event.Devices not returned, further analysis cannot be made.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history found no additional related issues for these items and the reported part and lot combinations.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Arcos Modular Revision Hip System
Common Device Name
Prosthesis, hip
Manufacturer (Section D)
Zimmer Biomet, Inc.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
Zimmer Biomet, Inc.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
+1(574)526-0384
MDR Report Key50193256
Report Number0001825034-2026-02318
Device Sequence Number20363958
Product Code KWA
UDI-Device Identifier00880304474215
UDI-Public(01)00880304474215(17)340216(10)66456848
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K090757
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number11-300820
Device Lot Number66456848
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2026
Initial Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
-
-