| Catalog Number |
11-300820 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Loss of Range of Motion (2032)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately two months post-implantation, the patient underwent a hip revision due to the patient being unable to internally rotate their foot.There was a suspected failure of the taper junction or proximal body locking screw.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).H6: proposed component code: mechanical (g04) ¿ stem.Complaint is not confirmed.Visual examination of the provided picture identified explanted stem and screw, bio-debris present, however it cannot be used to confirm the reported event.Devices not returned, further analysis cannot be made.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Review of complaint history found no additional related issues for these items and the reported part and lot combinations.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|