| Model Number |
471400-10 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted pulmonary lobectomy procedure, the long bipolar grasper instrument tip broke.A fragment fell into the patient and was retrieved during the same procedure.The customer confirmed the fragment was located in a suction canister.It was removed and matched with the long bipolar grasper instrument along the break line.The surgeon confirmed full fragment retrieval.No post-operative tests were performed.No additional procedure was required.The procedure was completed.There were no complications and the patient has not returned to the hospital due to experiencing any post-surgical complications related to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The long bipolar grasper instrument was analyzed and found to have a broken grip at the midpoint of grip.A piece approximately 2.85mm x 7.09mm was found to be broken off, confirming that a fragment detached from the long bipolar grasper instrument.The broken piece was not returned with the instrument.Further inspection of the returned long bipolar grasper instrument found no additional damage or abnormalities.
|
| |
|