| Model Number |
620-2231011-01 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device's screen turned completely white and was illegible.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.During investigation of the device to determine root cause, the technician observed physical damages to the front enclosure and top/bottom handle consistent with mis-handling.The front enclosure and top/bottom handle were all replaced to resolve the issue.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|