• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Trapezoid RX; lithotriptor, biliary mechanical

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Trapezoid RX; lithotriptor, biliary mechanical Back to Search Results
Model Number M00510890
Medical Device Problem Codes Break (1069); Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a trapezoid rx was to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2026.During preparation, the thumb ring detached from the handle.Another trapezoid rx was used to complete the procedure.There were no patient complications reported, and the patient's condition following procedure was stable.
 
Event or Problem Description
It was reported to boston scientific corporation that a trapezoid rx was to be used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2026.During preparation, the thumb ring detached from the handle.Another trapezoid rx was used to complete the procedure.There were no patient complications reported, and the patient's condition following procedure was stable.
 
Additional Manufacturer Narrative
Block h6: device code a0401 captures the reportable event of thumb ring break.
 
Additional Manufacturer Narrative
Block h6: device code a0401 captures the reportable event of thumb ring break.Block h11: no product analysis was performed because the complaint component was not returned for evaluation; therefore, the reported allegations cannot be confirmed.The device history record review confirmed that the device met all material, assembly, and performance specifications prior to distribution.A similar complaint review was completed.There were no emerging trends identified that warrant further action.Based on the information provided, the most probable cause of the reported issues "thumb ring break, basket failure to open and basket failure to close" cannot be established due to lack of evidence.The product was not returned for analysis to identify any defect with the device.Additionally, without proper evaluation of the device, it remains unknown the most probable causes that contributed to the event.Also, for the event additional device required, it happened during the procedure, and the most probable cause of this reported issue cannot be established due to lack of evidence.For this reason, will be classified as cause not established.All compiled information on this investigation determines that the most probable cause is cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Trapezoid RX
Common Device Name
lithotriptor, biliary mechanical
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
780 brookside drive
spencer IN 47460
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key50193693
Report Number3005099803-2026-03242
Device Sequence Number20361227
Product Code LQC
UDI-Device Identifier08714729296409
UDI-Public(01)08714729296409(17)261021(10)37758229
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K040447
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00510890
Device Catalogue NumberM00510890
Device Lot Number0037758229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2026
Supplement Date Received by Manufacturer09/04/2026
Initial Report FDA Received Date08/10/2026
Supplement Report FDA Received Date09/22/2026
Date Device Manufactured10/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexMale
Patient Weight171 LB
-
-