• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Philips Medizin Systeme Böblingen GmbH IntelliVue MX450 Patient Monitor; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Philips Medizin Systeme Böblingen GmbH IntelliVue MX450 Patient Monitor; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms) Back to Search Results
Model Number IntelliVue Patient Monitor MX450
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a complaint on an intellivue mx450 patient monitor indicating that the speaker malfunction.The device was not in clinical use at time of event, no adverse event was reported.The customer stated with the rse the speaker was not working.
 
Additional Manufacturer Narrative
E1; reporter institution phone number :(b)(6) e1: reporter phone number:(b)(6) analysis and testing performed by the philips field service engineer (fse) who went to the customer site.The issue was not replicated during the evaluation.As part of the troubleshooting process, all device connections were inspected, re-secured, and reset.Following reconnection of all applicable connections, the system was tested and found to be operating normally with no further issues observed.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause is unknown.Based on the available information, the issue appears to have been related to a connection-related condition that was resolved by re-establishing all connections.Normal system functionality was confirmed following the connection reset.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IntelliVue MX450 Patient Monitor
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
+49(703)14632031
MDR Report Key50194647
Report Number9610816-2026-101616
Device Sequence Number6111077
Product Code MHX
UDI-Device Identifier00884838038769
UDI-Public(01)00884838038769
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K130849
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIntelliVue Patient Monitor MX450
Device Catalogue Number866062
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2026
Initial Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/16/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-