• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Datascope Corp. Sensation Plus 8Fr. 50cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Datascope Corp. Sensation Plus 8Fr. 50cc IAB with Accessories, US Only; System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0576-01U
Medical Device Problem Codes Fluid/Blood Leak (1250); Difficult to Remove (1528)
Health Effect - Clinical Codes Hematoma (1884); Unspecified Vascular Problem (4441); Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: b4, g3, g6, h2, h3, h6 (type of investigation, investigation findings and investigation conclusions), h11.Corrected field: d4 (udi).The customer noted that patient-related factors such as age and vascular condition may have contributed to the event.Additional follow up indicated that the service representative was not requested to inspect or service the pump, and no blood back condition within the pump was reported.The device was retained by the customer for potential evaluation.A medical assessment was preformed and it concluded that the event is consistent with an iab membrane perforation, with therapy continuing despite indications of catheter perforation as outlined in the ifu.The resulting difficult catheter removal and vascular injury led to a hematoma and vessel repair; however, excessive force during removal cannot be confirmed as the cause.The reported complications are known risks already documented in the product risk file, and no new hazards were identified; therefore, no further escalation is required.The reported blood and clotting within the balloon and helium lumen are consistent with a potential balloon membrane breach, rupture, or fracture that may have allowed blood ingress into the catheter system.Difficulty during catheter removal and the subsequent hematoma could have been influenced by patient-specific factors, including vascular anatomy, vessel condition, procedural complexity, or tissue trauma associated with catheter placement and removal.Additionally, prolonged dwell time, vascular calcification, resistance encountered during removal, or mechanical stress on the balloon catheter may have contributed to the observed condition.Based on the available information, the exact cause cannot be conclusively determined without device evaluation.No decon or visual inspection performed since product was not returned and could not be evaluated.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.There were no ncmrs identified which could cause or contribute to the reported failure.The investigation does not indicate that the device was inadvertently released as non-conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend (increase in number of complaints over past three (3) months).The risk review was performed and no new risks, failure modes, causes or harms were identified.The trend review did not identify any adverse trends.No escalations are required at this time.Based on the rational provided above, no escalation to the capa process is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sensation Plus 8Fr. 50cc IAB with Accessories, US Only
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50194648
Report Number2248146-2026-0001386
Device Sequence Number6111078
Product Code DSP
UDI-Device Identifier10607567109633
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number0684-00-0576-01U
Device Catalogue Number0684-00-0576-01U
Device Lot Number3000525520
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2026
Supplement Date Received by Manufacturer08/10/2026
08/21/2026
09/15/2026
Initial Report FDA Received Date08/10/2026
Supplement Report FDA Received Date08/11/2026
09/08/2026
09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/02/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
Patient Age69 YR
Patient SexFemale
Patient Weight217 LB
-
-