• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Ethicon Endo-Surgery, LLC. Echelon Endopath; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Ethicon Endo-Surgery, LLC. Echelon Endopath; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number GST60B
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
T was reported that during an unknown procedure, it was observed that some staples fell off when they opened the packaging.Changed to a new one to complete the procedure.There was no patient consequence reported.No additional information can be provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 8/10/2026 d4: batch #985d34 investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the gst60b reload was received with no apparent damage.The sled was received in the home position; however, four double drivers and 2 single drivers were noted up without staples on the proximal end, this condition was most likely due to the sled movement.The condition of the driver's up shows that reload was partially fired 1/4 causing the reported event.However it is possible that the reload was manipulated, and the sled was manually returned to its home position.The returned reload was reset and loaded into a test device.It was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete, and the remaining staples met the staple form release criteria.The partially fired condition is consistent with an incomplete or interrupted cycle.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number of 109e93 / 22gst60b , and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 985d34, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Echelon Endopath
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR  00969
Manufacturer (Section G)
Ethicon Endo-Surgery, LLC.
475 calle c
guaynabo 00969
PR   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
PR   00969
+1(303)552-6892
MDR Report Key50195238
Report Number3005075853-2026-06451
Device Sequence Number18620707
Product Code GDW
UDI-Device Identifier10705036014706
UDI-Public(01)10705036014706(17)290430(10)109E93
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K183435
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberGST60B
Device Lot Number109E93
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/03/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2026
Initial Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-