| Model Number |
470056-08 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, a 30-degree endoscope plus was not responding to some commands.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) contacted the customer and obtained the following information: the endoscope rotates on its own ¿upside down¿ without the surgeon at the console using any commands.The failure was related to poor responsiveness.The movement of the surgeon did not scale appropriately to the endoscope.The movement was lesser.The endoscope moved in the intended direction and the issue happened during the procedure.The procedure was completed robotically and there was no injury to the patient.The procedure was prolonged by 15 - 30 minutes.No fragments fell into the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the 30-degree endoscope plus to perform failure analysis (fa).Fa was not able to confirm and reproduce the customer reported complaint.The endoscope was visually inspected and found with mechanical damage to the scope¿s distal tip.The root cause of is typically attributed to the use conditions, such as inadvertent collisions with hard surfaces or drop of the scope.
|
| |
|
Search Alerts/Recalls
|