| Catalog Number |
190858 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A user facility's biomedical technician (bmt) reported that a 2008t hemodialysis (hd) machine displayed a flow error in standby mode.A fresenius field service technician (fst) was called onsite and found multiple valves were corroded and the valve connections were melted.The fst replaced the balancing chamber assembly to resolve the issues.Machine functional tests were completed and passed onsite after repair.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The complaint device is not available to be returned to the manufacturer for evaluation.Multiple attempts have been made to contact the customer to obtain additional information related to the reported event.At this time, no additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|