| Model Number |
Patient Information Center iX |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on the patient information center ix indicating that the red alarms are acknowledging themselves, turning off without interaction, may not be latched.Missed an asystole on a patient last night, red alarm canceled itself.The device was in use on a patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A philips remote service engineer (rse) spoke with biomed, reviewed the alarm logs for the beds and confirmed that the reported red alarms were extreme tachycardia, not asystole.The logs showed these alarms were acknowledged by the cmu host and did not stop by themselves.The bedside configuration and alarm limits were checked and found to be appropriate.The rse explained to biomed that the alarms were working as configured and had been acknowledged from different hosts in the unit.Biomed confirmed understanding and reported that staff had acknowledged the alarms, with no harm to either patient.The system meets specification and was returned to use.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
| |
|