The customer returned the duodeno videoscope to the service center for a separate issue.During the device analysis, the investigation indicates that worn adhesive around objective lens, foreign objects inside of light guide lens and foreign material in forceps elevator was found.There was no patient involvement.
|
|
The device was returned to olympus for inspection and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: worn adhesive around objective lens, foreign objects inside of light guide lens and foreign material in forceps elevator.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: component failure.A root cause could not be identified.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material or root cause cannot be identified.A definitive root cause cannot be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|