It was reported that a patient stated that this inflatable penile prosthesis (ipp) exhibited a device deficiency.A computed tomography (ct) scan was performed, and the physician determined that the device needed to be replaced.During an evaluation, a possible sticky pump issue was identified, as the pump remained collapsed and the device did not inflate correctly.A surgical procedure was performed, during which a broken pump hose was identified.As a result, the ipp was removed and replaced.No patient complications were reported.
|