• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Infinion? CX; stimulator, spinal-cord, totally implanted for pain relief

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION Infinion? CX; stimulator, spinal-cord, totally implanted for pain relief Back to Search Results
Model Number SC-2317-70
Medical Device Problem Codes High impedance (1291); Unexpected Therapeutic Results (1631)
Health Effect - Clinical Code Pain (1994)
Date of Event 10/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was not getting pain relief due to high impedances on the spinal cord stimulation (scs) leads.Bsn sales representative was able to program around impedances, however patient opted to undergo a procedure wherein a full system was replaced.The patient was doing well postoperatively, and the explanted devices were disposed of.
 
Additional Manufacturer Narrative
Block d2b: pro code (product code): qrb.Additional suspect medical device component involved in the event: model number/catalog number: sc-1232.Serial number: (b)(6).Batch/lot number:558863.Model/catalog description: wavewriter alpha 32 ipg kit.Unique identifier (udi): (b)(4).Number/catalog number: sc-2317-70.Serial number: (b)(6).Batch/lot number: 7081053.Model/catalog description: infinion cx lead kit, 70cm.Unique identifier (udi): (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Infinion? CX
Common Device Name
stimulator, spinal-cord, totally implanted for pain relief
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key50196803
Report Number3006630150-2026-05719
Device Sequence Number10436338
Product Code LGW
UDI-Device Identifier08714729861638
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/15/2024
Device Model NumberSC-2317-70
Device Catalogue NumberSC-2317-70
Device Lot Number7080944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2026
Supplement Date Received by Manufacturer09/03/2026
Initial Report FDA Received Date08/10/2026
Supplement Report FDA Received Date09/24/2026
Date Device Manufactured12/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
-
-