• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIS Implants Technologies Ltd. C1 XD coni. con. implant D3.30 L13mm , NP; Implant, Endosseous, Root-Form

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIS Implants Technologies Ltd. C1 XD coni. con. implant D3.30 L13mm , NP; Implant, Endosseous, Root-Form Back to Search Results
Catalog Number C1-D13330
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient experienced dental implant loss.The implant was placed with less than 10 n of torque in fairly porous bone (type iv).
 
Additional Manufacturer Narrative
Since this event resulted in a serious injury, it is reportable per 21 cfr part 803.The device was not evaluated because the issue is a known inherent risk of the device.We will continue to track and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C1 XD coni. con. implant D3.30 L13mm , NP
Common Device Name
Implant, Endosseous, Root-Form
Manufacturer (Section D)
MIS Implants Technologies Ltd.
p.o. box 7
bar lev industrial park 20156 00
IS  2015600
Manufacturer (Section G)
MIS Implants Technologies Ltd.
p.o. box 7
bar lev industrial park 20156 00
IS   2015600
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
+1(717)845-7511
MDR Report Key50197179
Report Number3004203816-2026-19714
Device Sequence Number20207840
Product Code DZE
UDI-Device Identifier07290113996175
UDI-Public07290113996175
Combination Product (Y/N)N
Initial Reporter CountryDR
PMA/510(K) Number
K172505
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberC1-D13330
Device Lot NumberW25006729
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/21/2026
Initial Report FDA Received Date08/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexFemale
-
-