| Catalog Number |
NTLC75 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the scrub nurse properly loaded the cartridge to the body upon odp guide, but the surgeon was unable to fire the new tlc stapler all the way in order to transect the tissue.Was surgery delayed due to the reported event? no.Was procedure successfully completed? unknown.Were fragments generated? unknown.If yes, were they removed easily without additional intervention? unknown.Patient status/ outcome / consequences no.Was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 8/10/2026.D4 batch # 381d02.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ntlc75 device was received with no apparent damage and with a sr75 reload no loaded in the device.In addition, a tyvek was received along with the sample.The reload had the proximal 10 drivers up and the remaining drivers down with staples present and the swing tab in the locked position.Also, the gripping surfaces from right side was damaged marking the reload nonfunctional.The device was tested for functionality with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.It should be noted that the reload is designed to lockout, as a safety feature, if any staples have been fired from the cartridge reload.In addition, failure to complete the stroke may result in an incomplete staple line.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 381d02, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|