| Catalog Number |
CEV2295A |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A facility reported that the hook cev2295a 3pk n5 for hook handle (cev2295a) had a melted sheath.There was no impact to the patient; therefore, no injury, death, or surgical delay was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The hook cev2295a 3pk n5 for hook handle (cev2295a) was returned for evaluation: failure analysis: investigation findings showed that the blue coating melted at the end of the hook, revealing the metal of the hook.Root cause analysis: the blue coating melted, probably due to high voltage applied during use or contact with another instrument causing an electrical arc.This issue is being addressed in an ongoing corrective and preventive action (capa).
|
| |
|