INTUITIVE SURGICAL, INC None; ENDOSCOPE Back to Search Results
Model Number 470056-12
Medical Device Problem Code Image Orientation Incorrect (1305)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the image displayed on the 0 deg endoscope plus was upside down.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the site to obtain additional information: no further details could be provided regarding the reported event, other than the team reporting the image was inverted.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 0 degree endoscope plus was analyzed, and the findings did not replicate or confirm the customer reported complaint.The endoscope camera instrument adapter showed mechanical damage in the attached endoscope adapter (aea), including a stuck bearing gear.The endoscope cable integrated connector was visually inspected and found with mechanical damage.The cable integrated connector ferrule exhibited mechanical damages such as scratch marks, indentations or broken or missing pieces located at cable connector proximal end.The endoscope was visually inspected and found with minor cut(s) to the cable insulation.The endoscope was tested and placed on an in-house system for cable testing and failed due to wahoo paddleboard (wpb) defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
None
Common Device Name
ENDOSCOPE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50199512
Report Number2955842-2026-35166
Device Sequence Number5507565
Product Code GCJ
UDI-Device Identifier00886874116562
UDI-Public(01)00886874116562
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K191736
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470056-12
Device Catalogue Number470056
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/21/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/15/2026
Initial Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/06/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories
Patient SexUnknown
-
-