| Model Number |
470056-12 |
| Medical Device Problem Code |
Image Orientation Incorrect (1305)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the image displayed on the 0 deg endoscope plus was upside down.The procedure was completed with no reported injury.Intuitive surgical, inc.(isi) followed up with the site to obtain additional information: no further details could be provided regarding the reported event, other than the team reporting the image was inverted.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 0 degree endoscope plus was analyzed, and the findings did not replicate or confirm the customer reported complaint.The endoscope camera instrument adapter showed mechanical damage in the attached endoscope adapter (aea), including a stuck bearing gear.The endoscope cable integrated connector was visually inspected and found with mechanical damage.The cable integrated connector ferrule exhibited mechanical damages such as scratch marks, indentations or broken or missing pieces located at cable connector proximal end.The endoscope was visually inspected and found with minor cut(s) to the cable insulation.The endoscope was tested and placed on an in-house system for cable testing and failed due to wahoo paddleboard (wpb) defect.
|
| |
|