Institut Straumann AG SP ? 4.8mm RN, SLA? 6mm, TiZr, NTP; Endosseous Dental Implant Back to Search Results
Catalog Number 043.066S
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Fibrosis (3167)
Date of Event 07/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2026 in ada 19.Details of surgery: primary stability not achieved and implant surface not completely covered with bone.On (b)(6) 2026, non-osseointegration was verified.Patient presented with bone type iii and previous bone augmentation.The device was forwarded to the manufacturer.At the event the patient experienced: mobility.No further patient complications were reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.? the non-integration of an endosseous dental implant during the healing phase is a known inherent risk of the treatment with dental implants.Most implant failures occur before occlusal loading (analainen et al.2009).Based on clinical studies about 1-3% of the implants fail within the first year after implantation (ganeles et al.2008).Implants may have to be removed in case one or more of the implants success criteria are not met.Implant success criteria according to buser et al.(1991) are: 1.Absence of persistent subjective complaints such as pain, foreign body sensation and /or dysesthesia.2.Absence of a recurrent peri-implant infection with suppuration.3.Absence of implant mobility.4.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported early failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SP ? 4.8mm RN, SLA? 6mm, TiZr, NTP
Common Device Name
Endosseous Dental Implant
Manufacturer (Section D)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
Institut Straumann AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
+1(978)747-2509
MDR Report Key50199668
Report Number0009613348-2026-012421
Device Sequence Number13346546
Product Code DZE
UDI-Device Identifier07630031718532
UDI-Public07630031718532
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K150938
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number043.066S
Device Lot NumberXPMC3
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/15/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2026
Initial Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexMale
-
-