Catalog Number VCPB260H
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Holdl13.It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.The needle had been detached from the suture.It was not a control release needle.This event was found before use.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Holdl13.Product complaint # (b)(4).D4/g4: device is not distributed in the united states, but is similar to device marketed in the usa.Therefore, (01)gtin is not available.The following information was requested, but unavailable: - please provide the lot number.Unk.H3 investigation summary: the product was returned to ethicon for evaluation.One paper lid, one empty winding former, one detached needle and one dispensed suture were received for analysis.Product code vcpb260h.During the examination, a short suture insertion length was identified at the suture end.Additionally, no suture remnant was observed within the needle barrel hole.Subsequent visual inspection confirmed that the detachment was attributable to an inadequate suture insertion length, which did not conform to the specified acceptance criteria.Based on the information currently available, the short insertion issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Coated VICRYL
Common Device Name
Suture, Absorbable, Synthetic
Manufacturer (Section D)
Ethicon Inc.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
Ethicon Inc.
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
+1(303)552-6892
MDR Report Key50200498
Report Number2210968-2026-07947
Device Sequence Number5496491
Product Code GAM
UDI-Device Identifier10705031056343
UDI-Public10705031056343
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K181652
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCPB260H
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/09/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Initial Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-