Holdl13.Product complaint # (b)(4).D4/g4: device is not distributed in the united states, but is similar to device marketed in the usa.Therefore, (01)gtin is not available.The following information was requested, but unavailable:
- please provide the lot number.Unk.H3 investigation summary: the product was returned to ethicon for evaluation.One paper lid, one empty winding former, one detached needle and one dispensed suture were received for analysis.Product code vcpb260h.During the examination, a short suture insertion length was identified at the suture end.Additionally, no suture remnant was observed within the needle barrel hole.Subsequent visual inspection confirmed that the detachment was attributable to an inadequate suture insertion length, which did not conform to the specified acceptance criteria.Based on the information currently available, the short insertion issue was identified as pull out during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|