• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & Co. KG Working Element, bipolar

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & Co. KG Working Element, bipolar Back to Search Results
Model Number 26040EB
Medical Device Problem Code Unintended Electrical Shock (4018)
Health Effect - Clinical Code Electric Shock (2554)
Date of Event 05/22/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, following normal surgical procedures, hysteroscopic endometrial lesion resection and hysteroscopic diagnostic curettage were planned for the patient.During the use of high-frequency surgical equipment, the working handpiece suddenly malfunctioned and burned the bipolar electrode cable, causing transient electric shock to the patient.The surgical equipment was suspended immediately.After replacing the spare working handpiece and electrode cable and confirming no abnormalities, the operation resumed.Cross reference mdr 9610617-2026-01911.Cross reference mdr 9610617-2026-01912.
 
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not returned at the time of mdr submission.The event is filed under internal karl storz complaint id: (b)(4).Cross reference internal complaint id: (b)(4).Cross reference internal complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Working Element, bipolar
Common Device Name
Working Element, bipolar
Manufacturer (Section D)
KARL STORZ SE & Co. KG
dr.-karl-storz-strasse 34
tuttlingen 78532
MDR Report Key50201367
Report Number2020550-2026-01913
Device Sequence Number16036597
Product Code HIH
UDI-Device Identifier04048551091784
UDI-Public04048551091784
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/10/2026
Report Date (Section F) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26040EB
Device Catalogue Number26040EB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/30/2026
Event Location Hospital
Date Report to Manufacturer08/10/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/11/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UH400AUTOCON III 400 High End, 22Serial: (b)(6)
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-