| Model Number |
480440-06 |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/24/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci assisted surgical procedure, a white plastic tip (fragment) broke from the synchroseal instrument and fell into the patient.A fragment fell into the patient and was retrieved during the same procedure.Post operative x-ray has been performed to confirm the complete retrieval of the fragment.The procedure was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The synchroseal instrument was analyzed.Visual inspection found that the cut electrode was thermally damaged on the jaws.There was no material missing.This causes the jaws to pass energy inconsistently to the tissue.The synchroseal instrument was placed on in house system and passed engagement and recognition tests.The synchroseal instrument passed intuitive motion test.The jaws opened and closed properly.The synchroseal instrument could not be tested for sealing/cutting capabilities due to the instrument being returned with the cord cut.A review of the energy logs showed one "seal electrodes shorted".
|
| |
|