• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO Puerto Rico Mfg. Inc. PureSee IOL; Intraocular Lens

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO Puerto Rico Mfg. Inc. PureSee IOL; Intraocular Lens Back to Search Results
Model Number DET225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 06/24/2026
Type of Reportable Event Serious Injury
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PureSee IOL
Common Device Name
Intraocular Lens
Manufacturer (Section D)
AMO Puerto Rico Mfg. Inc.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50201612
Report Number3012236936-2026-000237
Device Sequence Number10274929
Product Code MJP
UDI-Device Identifier05050474819726
UDI-Public(01)05050474819726(17)290227
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberDET225
Device Catalogue NumberDET225U250
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/24/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/05/2026
Supplement Date Received by Manufacturer08/28/2026
Initial Report FDA Received Date08/11/2026
Supplement Report FDA Received Date09/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
Patient RaceWhite
-
-