• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Corentec Co, Ltd Bencox Hip System; Dual Offset RASP Handle (RIGHT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Corentec Co, Ltd Bencox Hip System; Dual Offset RASP Handle (RIGHT) Back to Search Results
Model Number HA.08A.0001
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 07/12/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was report that during surgery, the double offset broach handle repeatedly disconnected from the broach when attempting removal.Multiple broach handles were used and all disengaged from the broach.This caused the surgeon to apply reverse impaction at a variety of unusual angles in an attempt to remove the broach.During this reverse impaction, the femur was fractured.It is unknown if the femur was wired or not.*9 broach handles were used, therefore there are 9 separate reports for this event (report numbers: 2026-0001 through 2026-0009).
 
Additional Manufacturer Narrative
During surgery, the double offset broach handle repeatedly disconnected from the broach when attempting removal.Multiple broach handles were used, and all disengaged from the broach.This caused the surgeon to apply reverse impaction at a variety of unusual angles in an attempt to remove the broach.During this reverse impaction, the femur was fractured.It is unknown if the femur was wired or not.According to the customer complaint handling procedure (crsi-07-13),corentec shall attempt to retrieve the subject device from the hospital, surgeon, or facility so that a proper technical investigation can be conducted.While key product identification details (such as the production batch/lot number and serial number) have been verified, essential details regarding the nature of the issue including photos, videos, or the returned physical device have not been provided, making a complete root cause analysis unavailable at this time.A total of 4 units of the dual offset rasp handle (right) were affected across 3 lots during this same surgical procedure: 3 unit from lot 40ahhph0, 1 units from lot 40ahhpk3, and 1 unit from lot 20agkrh4.This mdr pertains to the 3 unit from lot 40ahhph0.Due to the inability to inspect the physical product, the root cause analysis cannot progress at this time.Corentec will continue post-market surveillance and attempt to obtain the necessary device details.Should any additional information be received that allows the investigation to resume, a supplemental mdr report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bencox Hip System
Common Device Name
Dual Offset RASP Handle (RIGHT)
Manufacturer (Section D)
Corentec Co, Ltd
12, yeongsanhong 1-gil
ipjang-myeon, seobuk-gu
cheonan-si, chungnam 31056
KS  31056
Manufacturer (Section G)
Corentec Co, Ltd
12, yeongsanhong 1-gil
ipjang-myeon, seobuk gu
cheonan si, chungnam 31056
KS   31056
Manufacturer Contact
jungsoo kim
12, yeongsanhong 1-gil,
ipjang-myeon, seobuk gu
cheonan si, chungnam 31056
KS   31056
MDR Report Key50201848
Report Number3009106092-2026-00004
Device Sequence Number10316574
Product Code HTQ
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Type of Report Initial
Report Date (Section B) 07/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHA.08A.0001
Device Lot Number40AHHPH0
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2026
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-