It was report that during surgery, the double offset broach handle repeatedly disconnected from the broach when attempting removal.Multiple broach handles were used and all disengaged from the broach.This caused the surgeon to apply reverse impaction at a variety of unusual angles in an attempt to remove the broach.During this reverse impaction, the femur was fractured.It is unknown if the femur was wired or not.*9 broach handles were used, therefore there are 9 separate reports for this event (report numbers: 2026-0001 through 2026-0009).
|
|
During surgery, the double offset broach handle repeatedly disconnected from the broach when attempting removal.Multiple broach handles were used, and all disengaged from the broach.This caused the surgeon to apply reverse impaction at a variety of unusual angles in an attempt to remove the broach.During this reverse impaction, the femur was fractured.It is unknown if the femur was wired or not.According to the customer complaint handling procedure (crsi-07-13),corentec shall attempt to retrieve the subject device from the hospital, surgeon, or facility so that a proper technical investigation can be conducted.While key product identification details (such as the production batch/lot number and serial number) have been verified, essential details regarding the nature of the issue including photos, videos, or the returned physical device have not been provided, making a complete root cause analysis unavailable at this time.A total of 4 units of the dual offset rasp handle (right) were affected across 3 lots during this same surgical procedure: 3 unit from lot 40ahhph0, 1 units from lot 40ahhpk3, and 1 unit from lot 20agkrh4.This mdr pertains to the 3 unit from lot 40ahhph0.Due to the inability to inspect the physical product, the root cause analysis cannot progress at this time.Corentec will continue post-market surveillance and attempt to obtain the necessary device details.Should any additional information be received that allows the investigation to resume, a supplemental mdr report will be filed accordingly.
|