• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS Pump; prosthesis, penis, inflatable

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CX MS Pump; prosthesis, penis, inflatable Back to Search Results
Model Number 72404282
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Impaired Healing (2378); Fluid Discharge (2686)
Date of Event 03/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this patient with an inflatable penile prosthesis (ipp) experienced continuous fluid discharge from the scrotal wound at the junction of the penis and the scrotum, and infection at the scrotum.The wound could not heal completely.A surgical procedure was performed during which the entire device was explanted.The patient was stable following the procedure, and no additional information was provided.
 
Event or Problem Description
It was reported that this patient with an inflatable penile prosthesis (ipp) experienced exudation of interstitial fluid from the scrotal wound, infection, and the wound could not heal completely.A surgical procedure was performed during which the entire device was explanted.The patient was stable following the procedure, and no additional information was provided.
 
Additional Manufacturer Narrative
Model number/catalog number: 72018201 serial number: n/a batch/lot number: 1100562589 model/catalog description: reservoir flat 100 ml unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.Infection, fluid discharge, and impaired healing are acknowledged within the ifu and are not related to off label use or failure to follow instructions.The reported patient symptoms are known risks associated with this device type and are indicated as such in the instructions for use.The device history record (dhr) confirmed that the device met all material, assembly and performance specifications.A risk review confirmed that the events of fluid discharge, healing impaired, and infection were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of adverse event related to procedure was assigned to this investigation as the patient had impaired healing and continuous fluid discharge from the site of the wound after the procedure from the infection that developed.Per the physician they described that before the procedure the patient showed no signs of infection, and the relevant tests were normal.However, within two to three after the procedure, there was repeated fluid leakage, which the physician considered to be an infection that occurred after the ipp implantation.No device malfunctions were reported.
 
Additional Manufacturer Narrative
Model number/catalog number: 72018201.Serial number: n/a.Batch/lot number: 1100562589.Model/catalog description: reservoir flat 100 ml.Unique identifier (udi) #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CX MS Pump
Common Device Name
prosthesis, penis, inflatable
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
+1(508)652-1217
MDR Report Key50202743
Report Number2124215-2026-42766
Device Sequence Number5507660
Product Code JCW
UDI-Device Identifier00878953003856
UDI-Public(01)00878953003856(17)290425(10)1100497541
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number72404282
Device Catalogue Number72404282
Device Lot Number1100497541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2026
Supplement Date Received by ManufacturerNot provided
09/01/2026
Initial Report FDA Received Date08/11/2026
Supplement Report FDA Received Date09/04/2026
09/21/2026
Date Device Manufactured04/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age31 YR
Patient SexMale
-
-