| Model Number |
SBC-501 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an orcapod 4 was used during a procedure performed on (b)(6), 2026.During the procedure, the middle section of the biopsy bung has detached and become lodged in the channel port, causing an obstruction in the working channel of the scopes.There were no patient complications reported as a result after this event.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Block d4, h4:
the complainant was unable to report the lot number; therefore the manufacture date and expiration date are unknown.Block h6:
imdrf device code a040103 captures the reportable event of seal biopsy valve material fragmentation.
|
| |
|