• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. 2008T HEMODIALYSIS SYSTEM W/BIBAG; Dialyzer, High Permeability With Or Without Sealed Dialysate System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. 2008T HEMODIALYSIS SYSTEM W/BIBAG; Dialyzer, High Permeability With Or Without Sealed Dialysate System Back to Search Results
Catalog Number 190766
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
A user facility biomedical technician (biomed) reported to fresenius technical support (ts) that a 2008t hemodialysis machine was experiencing an issue with its display screen.The biomed stated the screen was black.The biomed said they had already replaced the ui mics board, the ui to function board ribbon cable, and the screen.Ts advised the biomed to check the motherboard and all wired connections and connectors.Additional details were provided during follow-up with the biomed.Upon further investigation, the biomed identified thermal damage on the connector of the ground wire that goes from the power module to the main distribution board.The biomed replaced the whole power module (which included a new ground wire and connector), and this resolved the reported screen issue.No other components were found to be damaged.The distribution board did not have to be replaced.The biomed did not have any photos of the charred wire/connector.There was no burning smell noted, and there were no signs of any smoke, sparks, or flames.The biomed said the machine had 43,487 operating hours on it at the time of the event.The machine did not have any past problems failing the electrical leakage test, and it was plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.The biomed confirmed the machine was back in service and fully operational.There was no patient involvement associated with this event.The replaced parts were not available to be sent back for evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM W/BIBAG
Common Device Name
Dialyzer, High Permeability With Or Without Sealed Dialysate System
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
900 w zaragosa drive suite d
pharr TX 78577
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
+1(617)417-5172
MDR Report Key50202782
Report Number0008030665-2026-02158
Device Sequence Number5496559
Product Code KDI
UDI-Device Identifier00840861100910
UDI-Public00840861100910
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K231125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number190766
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Supplement Date Received by Manufacturer09/23/2026
Initial Report FDA Received Date08/11/2026
Supplement Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-