| Model Number |
471296-10 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the customer experienced an issue with an intuitive product.The customer reported event has no report of patient injury.Isi followed up with the initial reporter to obtain additional information; however, no further details are available regarding the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The large suturecut needle driver instrument was analyzed and found to have a frayed grip cable at the yaw pulley.Some wires were broken, but the cable itself was not fully broken.There was no discoloration, corrosion, or contamination observed on the cable that would occur from improper flushing or rinsing during reprocessing.Further inspection found no abnormally sharp or damaged components that could have contributed to the cable breaking.
|
| |
|