Model Number 380677-29
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, non-recuperable error occurred on the second surgeon side console (ssc).Logs show error 23008 was pointing to left master tool manipulator (mtm).The procedure was completing as planned with no reported injury.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: field service engineer (fse) confirmed issue was related to surgeon side console (ssc) 2 and it was disconnected.The procedure was completed using only ssc 1.The issue was not resolved over the phone, and the customer came to the conclusion to disable ssc 2.
 
Additional Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The fse was not able to reproduce the issue; however, the fse did replace the master tool manipulator (mtm).The system was tested and verified as ready for use.As of the date of this report, the mtm has not yet been received by intuitive surgical, inc.(isi) for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
daVinci Xi
Common Device Name
SURGEON SIDE CONSOLE, SMART PEDALS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50203281
Report Number2955842-2026-35167
Device Sequence Number5545397
Product Code NAY
UDI-Device Identifier00886874110744
UDI-Public(01)00886874110744
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K131861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number380677-29
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 07/14/2026
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/23/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories
Patient SexUnknown
-
-