A user facility biomedical technician (bmt) reported to technical support a 2008t machine powers down on its own randomly.There was no patient involvement.The bmt swapped the power switch, keypad assembly and the power logic board.The bmt was advised to swap the power supply (to isolate), possible power control board issue, functional board, check the motherboard for any issues, check all cables for problems.Upon follow-up, the bmt stated they were still troubleshooting the reported issue, and the machine was still out of service.The bmt mentioned that the power switch was burned, so it was replaced.It was confirmed there was no patient involvement, harm, or adverse events associated with the reported issue.Additional information was requested but was not received to date.
|