| Model Number |
394969 |
| Medical Device Problem Codes |
Premature Elective Replacement Indicator (1483); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
This device was explanted (b)(6) 2026.Patient presented to the emergency department on (b)(6) 2026.Per follow-up data showing changed parameters, the mode switch onset criterion was changed, and the ra sensitivity was decreased.The programming count changed from 22 to 25.It was also reported the patient underwent an mri on (b)(6) 2026, however there is not evidence that the device was properly put into mri mode.This device reached eri on (b)(6) 2026.Upon interrogation on (b)(6) 2026 it was reported an error message appeared.Device was explanted due to eri.
|
| |
|
Additional Manufacturer Narrative
|
|
Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|