| Model Number |
7731 |
| Medical Device Problem Codes |
Failure to Capture (1081); Fracture (1260); High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a generator replacement procedure, this right ventricular (rv) lead was determined to be fractured.Subsequent testing was performed, and loss of capture (loc) was observed.Additionally, pacing impedance measurements were found to be high and out of range.As a result, the rv lead was capped and a new lead was implanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|