| Model Number |
480460-12 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted distal gastrectomy surgical procedure, the sureform 60 stapler could not be removed due to the rubber cover breaking and blocking the entry into the trocar.The user was completing the procedure as planned with no further issue reported.No fragment fell inside the patient.No known impact or patient consequence was reported.Intuitive followed up with the initial reporter and obtained the following additional information: the instrument was inspected prior to use with no damage found.There was no instrument collision during the procedure.There was no fragment fall into the patient.The instrument was removed with the trocar.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did not receive the da vinci product with an alleged issue to perform failure analysis.
|
| |
|