Catalog Number 1758SI14
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use sterile water leak in the foley catheter.
 
Additional Manufacturer Narrative
The reported event is confirmed cause unknown.Evaluation determined that the product did not meet specifications.Visual: received 1 drainage bag with an inlet tube, sample port connector, catheter, 2 syringes, tes and packaging label.Verified material number 2758j14 and catheter lot number nglp3103.Functional: catheter filled with 10ml mbs using returned slip tip syringe.Upon inflation lumen to lumen leakage present.The labeling is found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions can be taken at this time.Correction: d,h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bardex® Lubri-Sil® I.C. All-Silicone Foley Catheter
Common Device Name
Silicone Foley Catheter
Manufacturer (Section D)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
+1(770)784-6100
MDR Report Key50206246
Report Number1018233-2026-05289
Device Sequence Number20201072
Product Code MJC
UDI-Device Identifier00801741025211
UDI-Public(01)00801741025211(17)280828(10)MYLR9450
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K984136
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1758SI14
Device Lot NumberNGLP3103
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2026
Supplement Date Received by Manufacturer09/21/2026
Initial Report FDA Received Date08/11/2026
Supplement Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-