The reported event is confirmed cause unknown.Evaluation determined that the product did not meet specifications.Visual: received 1 drainage bag with an inlet tube, sample port connector, catheter, 2 syringes, tes and packaging label.Verified material number 2758j14 and catheter lot number nglp3103.Functional: catheter filled with 10ml mbs using returned slip tip syringe.Upon inflation lumen to lumen leakage present.The labeling is found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Based on the results of the investigation, no additional actions can be taken at this time.Correction: d,h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|