Investigation:
the patient, whose clinical signs and symptoms are unknown, had a blood culture sample collected on (b)(6) 2026.Initially, the sample was tested at the customer¿s affiliate site on (b)(6) 2026 with the bcid2 panel, which reported methicillin-resistant staphylococcus aureus (mrsa) as detected.On (b)(6) 2026 at the customer's site, the patient¿s blood culture sample was tested on the biofire® blood culture identification 2 (bcid2) panel.The bcid2 panel reported mrsa as detected.The result reported to the physician was meca/c and mrej detected.On an unspecified date, additional testing was performed.Gram stain yielded gram positive cocci in clusters.The isolated organism was identified via latex coagulase, mrsa screen plate, and pbp2a antigen kit.Susceptibility testing was performed on vitek ast-gp75.Susceptibility testing indicated that the organism was methicillin-susceptible.The customer reported that due to the bcid2 panel results, the patient was put on vancomycin and ceftriaxone.Vancomycin was discontinued after methicillin sensitive staphylococcus aureus (mssa) was confirmed.Repeat blood cultures from (b)(6) 2026 were negative and the patient was stepped down to a skilled nursing facility (snf).No patient harm was reported.The analysis of the bcid2 panel run files provided by the customer showed late melt signatures for mrej, suggesting that the target nucleic acid was present at low levels.The run files also showed moderate melt signatures for meca/c, suggesting that the target nucleic acid was present at a level that would be expected to be reproducible.Qc records for pouch/kit lot 3zjr26/0643126 and 3xl725/2504225 and instruments tm05669 and tm19280 were reviewed.All qc metrics for the pouch/kit lots and instruments were met and passed.Instruments appeared to be operating as intended.Conclusion:
the investigation determined that the most likely cause of the discrepant methicillin-resistant staphylococcus aureus (mrsa) result was sensitivity/specificity differences between the bcid2 panel and comparator phenotypic susceptibility methods.The bcid2 panel is a multiplexed nucleic acid test intended for simultaneous qualitative detection and identification of multiple bacterial and yeast nucleic acids and select genetic determinants associated with antimicrobial resistance.While rare, discrepancies between biofire® and other laboratory methods are part of the normal performance of the product observed in the field.According to the biofire® blood culture identification 2 (bcid2) panel instruction booklet, it is possible to obtain a "detected" result for staphylococcus aureus meca/c and mrej (mrsa) using the bcid2 panel but to recover an isolate from culture that is characterized as methicillin-sensitive s.Aureus (mssa) using phenotypic ast methods.This may be observed in instances of heterogeneous cultures of mrsa and mssa.This occurs when a population of resistant and susceptible s.Aureus are present within the clinical sample, and standard culture methods are unable to isolate/differentiate a resistant population of s.Aureus.Please note that, per the bcid2 ifu, the results for the antimicrobial resistance gene assays do not specifically link the resistance gene to the applicable bacteria detected.Antimicrobial resistance can occur via multiple mechanisms.A detected result for a genetic marker of antimicrobial resistance cannot be definitively linked to the microorganism(s) detected.Culture is required to obtain isolates for antimicrobial susceptibility testing, and bcid2 panel results should be used in conjunction with culture results for the determination of susceptibility or resistance.Due to the high sensitivity/specificity of the bcid2 panel and the ability to detect organisms/antimicrobial resistance genes directly from the patient's blood culture rather than isolated culture colonies, discrepancies with standard identification techniques can occur and are a part of normal system performance.False positive results can be the result of a variety of sources and causes.Detection of organism targets does not imply that the corresponding organisms are infectious or the causative agents for clinical symptoms.Results from this test are intended to be correlated with the clinical history, epidemiological data, and other data available to the clinician evaluating the patient.All identification results provided by the bcid2 panel are intended to be interpreted in conjunction with gram stain results.Clinical performance can be found in table 81 of the biofire® blood culture identification 2 (bcid2) panel instruction booklet: https://www.Biofiredx.Qarad.Eifu.Online/iti/all?keycode=iti0048.
|
|
(b)(6) ((b)(6)) reported potential false positive methicillin-resistant staphylococcus aureus (mrsa) results on the biofire® blood culture identification 2 (bcid2) panel after testing a patient blood culture sample.The customer reported that due to the bcid2 panel result, that patient experienced indirect harm, specifically the patient was put on vancomycin and ceftriaxone, which might lead to serious injury if it were to recur.No patient harm was reported.The investigation determined that the most likely cause of the discrepant mrsa result was sensitivity/specificity differences between the biofire® bcid2 panel and comparator phenotypic susceptibility methods.
|