| Model Number |
48345B-03 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted surgical procedure, the sureform 45 blue reload was found to be damaged.After the procedure, fragments consistent with the reported damage were discovered in a saline bottle used for post-procedure cleaning by the central sterile processing department.A backup da vinci reload was used to complete the procedure.The procedure was completed as planned with no reported patient injury.Intuitive surgical, inc.(isi) contacted the customer and obtained the following information: there was no damage/no patient impact was reported.(the fragment was discovered during the procedure and cross-checked against the cartridge.) the stapler was used as-is, and the operation was completed.As a precaution, all used reload units have been sent to you.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the sureform 45 blue reload to perform failure analysis (fa).Fa was able to confirm and reproduce the customer reported complaint.The reload was found to have exterior damage at one of the proximal flanges of the cartridge.There may be missing material, but the size of the missing pieces cannot be confirmed.The probable root cause is contributed to use conditions (i.E.Due to reload removal).Additional observation not reported by site: the reload was found to have undeployed pusher.One of the pushers was undeployed with "b" shaped staple stuck in the staple pocket near the distal end of the reload.The pusher was manually activated and without much effort it moved to its final position, pushing the staple out of its pocket.The staple was lost during fa.
|
| |
|
Search Alerts/Recalls
|