(b)(4).Date sent 8/11/2026.D4 batch # unk.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.D4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.B3: only event year known: 2026.An analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.Additional information was requested, and the following was obtained:
what is the product code? it is believed to be either cdh25p or cdh29p.What were the indications for surgery? unsure of question.What surgical procedure was performed? sigmoid colon resection.Did the patient receive any preoperative chemotherapy or radiation? no low.Were there any issues experienced with the device in the initial surgical procedure? no.What healthcare professional fired the device and what is his/her experience with the device? been a user since the device came to market.Where in the green gap setting scale was the inoicator located prior to firing (low-b, middle-b, or high-b)? low b.Did the healthcare professional wait 15 seconds aftyes, losing the device and then retighten prior to firing? yes.Were there any issues with device use/firing? no.What confirmation was received that the device completed the firing sequence? green check.Was the green checkmark visible at the end of the firing? yes.How many counter-clockwise revolutions of the adjusting knob were used to open the device? 2.Was there any difficulty removing the device? no.Was a complete transection of the white breakaway washer visually confirmed? yes.Were the donuts inspected? if so, please describe.Yes, two complete donuts.Were there any issues noted with staple formation? if so, please describe the shape and location.Not observed.How many days postoperative did the leak occur? unknown.How was the leak identified? infection.What was observed at the site of the leak upon reoperation? unknown.How was the leak addressed? oversew.What is the current status of the patient? unknown.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|