|
A healthcare professional reported that during a cerebral arteriovenous malformation embolization, an
envoy da, 6f, 105cm, mpd guide catheter (product code: 67125805d, lot number: 31892614) was in place and a competitor¿s microcatheter (mc) became impeded in the middle section of the guide catheter and could not advance any more.The doctor only retracted the microcatheter and switched to a second competitor¿s microcatheter.The second microcatheter had the same issue.The doctor removed the guide catheter and the second microcatheter from the patient together, then switched to a new guide catheter to complete surgery with the second microcatheter.Additional event information received on 31-jul-2026 indicated that they were able to move the device.They were not able to torque the device.There was no evidence of physical material within the device.There was no difficulty removing the device from the patient.There was no procedure delayed/prolonged due to the event.
|
|
Manufacturer ref#(b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section e1.Initial reporter phone: (b)(6)
an analysis of the product could not be performed since a physical sample was not received for evaluation.A manufacturing record evaluation was performed for the finished device 31892614 number, and no internal actions related to the malfunction were identified.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no internal actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|