| Model Number |
02.12.E0410FL |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 11 month from the primary, the patient came in due to signs of infection and the pathogen is unknown.The surgeon performed a washout and revised the flex4l - 10mm insert to a same size insert.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 07 august 2026 gmk-spherika 02.12.E0410fl tibial insert e-cross - flex 4l - 10mm () lot 2507483: (b)(4) items manufactured and released on 14-apr-2025.Expiration date: 31-mar-2030.No anomalies found related to the problem.To date, 99 items of the same lot have been sold with no similar reported event during the period of review.Root cause: infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|