| Model Number |
02.07.1204L |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Medical device site infection (4845)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 1 year and 9 months from the primary, the patient came in due to signs of infection and the pathogen is unknown.The surgeon performed a washout and revised the gmk-sphere cemented s4 tibial tray to a rev.Tibial tray left s4 and revised the flex 4l - 11mm insert to a flex 4l - 13mm at his discretion.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 06 august 2026 gmk-sphere 02.07.1204l tibial tray fix cemented s.4l () lot 2348228: (b)(4) items manufactured and released on 02-may-2024.Expiration date: 15-apr-2029.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Gmk-sphere 02.12.E0411fl gmk-sphere tibial insert e-cross - flex 4l - 11mm () lot 2116927: (b)(4) items manufactured and released on 12-may-2022.Expiration date: 19-dec-2026.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review root cause:infection is a known possible complication of any surgery.Although no root cause can be established, there is no indication that any issue with the device may have caused or contributed to the event, and the document review does not indicate any potential manufacturing related cause.
|
| |
|