| Model Number |
72404132 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with an artificial urinary sphincter (aus) experienced recurrent urinary incontinence.It was determined that the cuff was too large for the patient.A surgical procedure was performed to downsize the cuff.No further patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with an artificial urinary sphincter (aus) experienced recurrent urinary incontinence.It was determined that the cuff was too large for the patient.A surgical procedure was performed to downsize the cuff.No further patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 1100701246.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100720950.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 1100701246.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100720950.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the event of urinary incontinence and inadequate cuff size/loose cuff was defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available the cause that contributed to the reported event cannot be established.Therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|