• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries,LP MEDLINE; PACK,COLD,INSTANT,STANDARD,6"X9"

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries,LP MEDLINE; PACK,COLD,INSTANT,STANDARD,6"X9" Back to Search Results
Catalog Number MDS137000
Medical Device Problem Code Excessive Cooling (2932)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the cold pack she was using caused a burn on her neck.Customer noticed a couple hours after using it has blistered.
 
Additional Manufacturer Narrative
It was reported that ¿customer states they cold pack she was using caused a burn on her neck.Customer noticed a couple hours after using it has blistered.¿ there were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
PACK,COLD,INSTANT,STANDARD,6"X9"
Manufacturer (Section D)
Medline Industries,LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
arundathi nomula
three lakes drive
northfield, IL 60093--275
MDR Report Key50209714
Report Number1417592-2026-15367
Device Sequence Number15952219
Product Code IMD
UDI-Device Identifier40080196732334
UDI-Public40080196732334
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 07/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS137000
Device Lot Number48823090001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2024
Initial Report FDA Received Date08/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/01/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
Patient SexFemale
Patient Weight200 LB
-
-