A review of the complaint history for (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for (b)(6) was performed from the date of manufacture, 12-apr-2021 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, a technical support specialist found that customers did not have cycle count privileges, reviewed the populate buttons screen and found no failed drawers, checked the windows event logs and found no relevant errors related to the reported issue, reviewed med bank my quick link (mql) transactions and confirmed that a single tablet had recently been issued to the patient from bin 02 01, assisted customer with performing a cycle count on bin 02 01 for user, which allowed retrieval of the tablet that had not been accessible previously.Issue was resolved, and normal functionality was restored.The system functioned as intended after the technical support specialist investigated the device.
|