C.R. Bard Inc. (Covington) -1018233 Bardex® I.C. Foley Catheter Back to Search Results
Catalog Number 0165SI14
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during placement during surgery when urine volume was checked during surgery, there was almost no urine accumulated.The foley catheter had spontaneously dislodged, and the leak location was unknown.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing related.The reported failure was able to be reproduced.The product was used for urological care.Based on the evaluation, a tear was found on the rubberized layer causing interlumen leakage.Hence, this complaint was confirmed related to manufacturing process.A potential root cause for this failure could be insufficient latex strength, low/non-uniform rubberize thickness, incorrect wire pressing technique, incorrect stripping technique etc.No deviations or nonconformances related to the reported issue were identified.The labeling / packaging record review did not identify any labeling / packaging related conditions associated with the reported failure mode.Manufacturing process indicates that the process is capable of producing of this type of defect.However, a review of the device history record did not show any problems or condition that would have contributed to the reported issue.Current in process controls per pfmea are adequate to identify the defect or verify the product integrity.Therefore, no additional action required at this time.Corrections: d,e,h.Udi format updated with available product information per gudid.H11: section a through f, the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Bardex® I.C. Foley Catheter
Common Device Name
Foley Catheter
Manufacturer (Section D)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. Bard Inc. (Covington) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
+1(770)784-6100
MDR Report Key50209941
Report Number1018233-2026-05306
Device Sequence Number15952102
Product Code MJC
UDI-Device Identifier00801741016745
UDI-Public(01)00801741016745(17)270728(10)MYLQ2826
Combination Product (Y/N)Y
Initial Reporter StateNA
Initial Reporter CountryJA
PMA/510(K) Number
K910318
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0165SI14
Device Lot NumberMYLQ2826
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/18/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2026
Supplement Date Received by Manufacturer09/27/2026
Initial Report FDA Received Date08/11/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-